Text here....
|
UCD-60 - new universal industrial ultrasonic flaw detector
 | |
New universal ultrasonic flaw detector allowing to take advantage of all riches opportunities of modern digital techniques: to deduce a signal in the form A, B, C-scan, to connect the transducer on way for construction coordinate range of a product, to write down great volume of data, automatically to form reports and to print them directly from the device. Allows to measure thickness of products with high accuracy and correction of V-figurativeness, to spend manual, automated contact and immersion (special and-zone) the testing.
Allows connecting multichannel scanning systems (up to 32 channels).
Certificated of Gosstandart of Russia RU.C.27.003. A №27905 is registered in the National register of means of measurements under №34808-07 also admitted to application in the Russian Federation.
|
- Display: color TFT 135 x 100 mm (640 х 480 pixels)
- Amplifier: 0.4 - 20 МHz
- TGC/DAC functions: 90 dB, 12dB/mks
- Work with curve amplitude-interval: measurement in dB echelon signal relative curve and operation LED
- Pulse repetition frequency: to 2000Hz
- Measurement signal spectrum
|
- Monitor gate: three independent monitor gates with individual logic sound indication
- А-, В-, С- scan
- Possibility to connect the transducer of a way
- Connection multichannel scanning system (to 32 channel)
- Interface RS232, Ethernet
- Mains power: LiOn batteries or via external plug-in power unit
- 7 hours battery operation
|
Basic Delivery Set
- UCD-60 main unit
- AC-DC power supply 220 V
- Probe cable SR50-Lemo 2 pieces
- Four transducers
- Interface cable PC
- Software support for connection PC
- Documentations
- Transport bag
Additional implements
- Protective bag
- Batteries
- Ultrasonic transducers for different application
- Standard samples
- Scan mechanism
- Multiplexer
|
Technical Characteristics
- Calibration range
- min.: 0 - 2,67mks
max.: 0 - 1000mks
0,01 / 0,1/ 1/ 10/ 100mKs - Pulse shift
- -4mks to 1000mks
0,01 / 0,1/ 1/ 10/ 100mks - Maximal length controllable material
- 6000 mm
- Sound velocity range
- 1000 - 9999 m/s
- Probe delay
- 0 - 100 mks
with step 0,01 / 0,1/ 1 and 10mks - Damping
- 25 ohms / 50 ohms/ 1000 ohms
- Resiver
- 50 оhms/ 600 ohms
- Probe pulse
- radio-frequency pulse, amplitude 50 and 200 V.
- Damping probe pulse
- controllable up to 0 to 15 half-period
with delay damping up to 0 to 7 half-period - Pulse repetition frequency
- 20 to 2000Hz
1/ 5/ 10 or 100Hz - Amplifier
- 0.4-20 MHz (-6 dB)
- Gain
- 100 dB, adjustable in steps 0.5, 1, 2 or 6 dB
- TGC/DAC
- 70 dB, 12 dB/mкs
- Rectification
- RF (in all ranges), full-wave,
full-wave, positive and negative half wave B-scan, C-scan - Reject
- 0 - 90% screen height
- Monitor gate
- three independent monitor gates with individual logic and sound indication
- Alarm
- luminous and sonorous
- Measurement resolution
- digital display of sound path between zero and the first echo signal
- Display of amplitude
- as a percentage of the screen height,
in dB to reference test echo,
in dB reference to DAC curve. - Display
- color, TFT 640 х 480 pixels
130 х 100 mm - А-signal
- 480 x 300 pixels in basic mode
- Memory
- 500 datasets with с А-signal
5000 checking protocol(signal, envelope, measurement result, working conditions, date, time and name protocol) - Interface
- RS232 / Ethernet
- Dimensions transducers
- 2 SP50 or Lemo0 (under order)
- Batteries
- LiOn 8 А/hr
- Operating time
- 7-8 hours from batteries
- Mains power
- 220 V AC
- Operating temperature
- up to -20 C to +50 C
- Dimensions (H x Wx D)
- 190 mm x 285 mm x 50 mm
- Weight
- 3 kg with batteries
|
|
|