Text here....
|
UDT-40- precision thickness gauge with A-scan
 | | Precision, not having analogues of Russia, universal ultrasonic thickness gauge with high reliability of the testing. Presence A –scan excludes such typical mistakes at measurement of thickness as doubling of indications, and B-scan allows observing a structure of a bottom of a product. Thickness gauge realizes various ways of the testing of thickness - measurement of time on transition through "zero", measurement of time between signals, and also frequency and correlation methods of measurement.
Presence of the built in contours of the coordination, adjustable width of a pulse and three electric damping allows to use for measurement any transducers with frequencies from 0,5 up to 15 MHz, both single element and dual type.
|
- Range: 3 to 2975 mm
- Display: LCD/ELD 240 х 128
- Amplifier: broadband 05-15 MHz
- Connection any modern transducers
- Controllable gain: 90 dB
- Fixing a minimal unit value gauge
- A-scan, B-scan(mud line products
- "Freeze" А-signal
- Correction V-imagery transmission transducers
|
- Measurement gauge one after one signal and between two signal
- Precision indication gauge: 0.001, 0.01 and 0.1
- Memory configuration (all parameters device and transducers)
- Memory results with possibility watch on-screen
- Interface RS232
- Mains power: 4 Nimh batteries size "D" or external block
- Minimum 8 hours from batteries
- Weight 2 kg with batteries
|
Basic Delivery Set
- UDT-40 thickens gauge electronic module
- Electric power supply 220 V
- Probe cable SR 50-Lemo 2 pieces
- Two transducers
- Interface cable RS 233 for connection PC
- Software support for PC
- Transport bag
Universal Delivery Set
- Base delivery set
- Protective boot
- Module for batteries size "D" 9000
- Set batteries NiMh size "D" 9000 mA/hr
- Universal charging device
Additional implements
- Adapter circuit for connection to USB to port of the PC
- Sample of gauge "stupenika"
- Ultrasonic transducers for diverse application
- Certificated sample of gauge from different fusion
|
Technical Characteristics
- Calibrition range
- min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks - Sound velocity range
- 1000 - 9999 m/s
- Pulse shift
- up to -0,5 mks to 996 mks
- Damping
- 50 оhms / 1000 ohms (to 25 ohms in simultaneous mode)
- Input impedance
- 50 оhms / 600 оhms
- Sound impulse
- quadrangular, amplitude 200V,
with change interval up to 50 to 500 ns,
with step 25 ns - Amplifier
- broadband 0.5-15 MHz (-6 dB)
- Controllable gain
- 90 dB, with step 0.5, 1, 2 and 6 dB
- Rectification
- positive or negative half wave,
complete, radiofrequency signal - Reject
- compensate, 0-80% screen height
- Inspection zone
- two independent monitor gates with individual logic
- Alarm
- luminous and sonorous
- Measurement resolution
- to 0 to first signal in zone or between
signals in zones - Dispaly
- LCD with regulate contrast and backlight
or with regulate brightness
110х65 mm; 240х128 pixels
- Memory
- 100 datasets for transducers
750 results of measurements with A-scan, parameters of the transducer, date and time - Interface
- RS232 or USB
- Size transducer
- Lemo0 or BNC (optional)
- Batteries
- 4 NiMh size "D"
- Operating time
- minimum 8 hours
- Mains power
- external block power 220V AC
- Power consumption
- maximum 5W, depending
on setting parameters - Operating temperature
- up to -20 C to +55 C (ELD)
up to -5 С to +55 С (LCD) - Dimensions (H x W x D)
- 160 mm x 225 mm x 45 mm
- Weight
- 2 kg with 4 batteries size "С"
|
|
|